AQX/P
The acquisition control system AQX/P generates
all digital control signals as specified by the user’s pulse program.
The CCU runs as a "diskless UNIX" client and controls the TCU, 2
FCU’s, RCU and GCU. The TCU controls all timing and pulses. The
FCU’s generate RF phases, frequencies, and control the RF amplitudes.
Amplified, demodulated and digitized NMR signals are fed to the
RCU which performs real time digital filtering and accumulation.
The optional GCU generates the gradient shapes for GRASP experiments. |
AQR/P
The acquisition rack AQR/P contains all
the RF/ AF components for signal routing, transmitting and receiving
purposes. The NMR signal is mixed down to audio frequency by the
RX22 receiver system and then converted to digital data by the ADC.
The local oscillator frequency is derived from the ASU/LOT module. |
BSMS
The BSMS unit provides NMR sample control
(SLCB), shim (SCB) and H0 current (LCB) control, houses the digital
lock (LTX, LRX) and the optional gradient amplifier (GAB). |
HPPR preamplifier module
The HPPR amplifies, filters and routes
the NMR response signals from the probe to the RX22 receiver. It
switches the RF transmitter output to the probe. |