informatica:hp_xw4400_400sb_topspin31 [2014/02/19 12:27]
miquel [Afegir sermnadmin al grup 'nmrsu']
informatica:hp_xw4400_400sb_topspin31 [2014/06/26 11:34]
miquel [Problemes amb 'cf']
Line 879: Line 879:
 chmod 775 * chmod 775 *
 chmod 664 *i chmod 664 *i
-chown root:root acq_control nmrsim.mod shimcntl xwish3+chmod u+s xwish3 
 +chown root:root xwish3
 </code> </code>
  
Line 894: Line 895:
  
 root@cie-55-31:/opt/topspin3.1/prog/mod# ls -l | grep root root@cie-55-31:/opt/topspin3.1/prog/mod# ls -l | grep root
--rwxrwxr-x 1 root  root   3927964 Feb 18 16:53 acq_control +-rwsrwxr-x 1 root  root   6913656 Feb 12  2012 xwish3
--rwxrwxr-x 3 root  root   2852204 Feb 12  2012 nmrsim.mod +
--rwxrwxr-x 1 root  root   2984972 Feb 18 16:39 shimcntl +
--rwxrwxr-x 1 root  root   6913656 Feb 12  2012 xwish3+
 </file> </file>
  
Line 910: Line 908:
 Done. Done.
 </code> </code>
 +
 +===== L'amplificador de gradients no funciona =====
 +
 +L'amplificador de gradients no funciona. Al procés de configuració del hardware es detecta correctament, però a l'hora d'adquirir els experiments o quan es fa servir el //topshim// no fa res. Després de provar a reiniciar-ho, aturar-ho, i desconnectar-ho sense resultats, faig una consulta a n'Òscar i resulta que la solució es canviar el valor del paràmetre ''GRADRES'' al fitxer de configuració //uxnmr.par//,
 +
 +<file>
 +Subject: Re: Error quan faig CF amb Topspin-3 al 600 i 400sb
 +Date: 2014-02-19 15:06
 +From: Óscar Benítez <...@bruker.es>
 +To: Miquel Cabanas <...@uab.cat>
 +
 +Mira el parametre GRADRES del fitxer /opt/Topspin3.x/conf/instr/spect/uxnmr.par
 +
 +Hauria de ser 5. Si no ho és canvia'l.
 +
 +Oscar Benítez Castells
 +NMR Service Engineer
 +</file>
 +
 +Comprovo que el valor actual no és el correcte, el modifico i l'amplificador de gradients comença a funcionar.
  
 ====== Adreces IP dels mòduls de la consola ====== ====== Adreces IP dels mòduls de la consola ======
Line 921: Line 939:
   * [[http://149.236.99.91 | Bruker Gradient Amplifier - Device Information]]   * [[http://149.236.99.91 | Bruker Gradient Amplifier - Device Information]]
   * [[http://149.236.99.20 | BSMS Service Web - Main Menu]]   * [[http://149.236.99.20 | BSMS Service Web - Main Menu]]
 +
 +====== Problemes amb 'cf' ======
 +
 +Cada cop que s'executa la comanda ''cf'', surten en pantalla uns missatges d'avís per indicar que el programa no ha pogut crear uns enllaços (//hardlinks//) necessaris pel funcionament del Topspin i que caldrà crear-los a mà. Per exemple,
 +
 +<code>
 +hconfserver
 +
 +Operation not permitted
 +Cannot make link
 +link must be done manually with
 +  ln /opt/topspin3.1/prog/mod/go4 /opt/topspin3.1/prog/mod/acq_control
 +
 +Operation not permitted
 +Cannot make link
 +link must be done manually with
 +  ln /opt/topspin3.1/prog/wobble/pp_birds /opt/topspin3.1/prog/wobble/pulsprog_X
 +</code>
 +
 +Per resoldre provisionalment el problema,
 +  * he escrit un petit //script// amb les comandes a executar per crear els enllaços i copiar un fitxer (''shimcntl'') que de vegades desapareix quan s'executa la comanda ''cf''; i
 +  * he configurat el sistema perquè l'usuari //sermnuab// pugui executar aquest //script// amb els permisos necessaris.
 +
 +<WRAP center round tip 60%>
 +**Després d'executar la comanda 'cf' al Topspin,**
 +  * obre una finestra per comandes, i
 +  * escriu la següent ordre:\\ ''sudo /home/sermnuab/bin/fix_topspin31.sh''
 +</WRAP>
 +
 +
 +
  
  
  
  