This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
informatica:hp_z400_600_topspin363_2 [2021/08/30 11:49] miquel [PENDENT: Configuració del MICS] |
informatica:hp_z400_600_topspin363_2 [2021/08/31 10:34] (current) miquel [PENDENT: Instal·lació del gestor de llicències CodeMeter] |
||
---|---|---|---|
Line 288: | Line 288: | ||
===== PENDENT: Configuració del MICS ===== | ===== PENDENT: Configuració del MICS ===== | ||
+ | |||
+ | Comprovo que MICS ha creat els fitxers '' | ||
<WRAP todo> | <WRAP todo> | ||
- | | + | Caldrà afegir/ |
- | * Descarregar-se el fitxer BIS de l' | + | |
- | * [[http:// | + | |
+ | * cryoshims anterior | ||
+ | * registre | ||
+ | * registre de recàrregues de LHE | ||
+ | * registre de recàrregues de LIN | ||
</ | </ | ||
Line 322: | Line 329: | ||
com que no se'ns aplica, premo el botó < | com que no se'ns aplica, premo el botó < | ||
- | ^ Magnet | + | Llavors s'obre un nou formulari amb un avís, |
- | | Magnet BZH Number | + | |
- | | Frequency | + | |
- | | Cryobore | + | |
- | | Design Letter | + | |
- | | Design Version | + | |
- | ^ Cryostat | + | |
- | | Cryostat Serial | + | |
- | | Dewar Type | + | |
- | | RT Bore | 54 mm | | + | |
- | ^ Helium | + | |
- | | He Total Volume | + | |
- | | He Holdtime | + | |
- | | He Boil-Off | + | |
- | ^ Nitrogen | + | |
- | | N2 Total Volume | + | |
- | | N2 Holdtime | + | |
- | | N2 Boil-Off | + | |
- | | N2 Vessel Height | + | |
- | ^ Helium Sensor | + | |
- | | Part Number | + | |
- | | Min. Allowed Level | 46 % | | + | |
- | | Min. Level Energizing | + | |
- | | Active Length | + | |
- | ^ He level graph ^^ | + | |
- | | Point1 Level | + | |
- | | Point1 Volume | + | |
- | | Point1 Level | + | |
- | | Point1 Volume | + | |
- | | Point1 Level | + | |
- | | Point1 Volume | + | |
- | ^ Helium Sensor (if present) | + | |
- | | Part Number | + | |
- | | Min. Allowed Level | % | | + | |
- | | Active Length | + | |
- | + | ||
- | + | ||
- | + | ||
+ | < | ||
+ | BIS Parameter Definition | ||
+ | This form must be used for NMR magnets only. | ||
+ | Please do this very carefully. | ||
+ | Parameters should be entered based on magnet manual and type label. Entering wrong data might result in erroneous indications and/or false alarms in system monitoring. | ||
+ | </ | ||
+ | Després de llegir l' | ||
+ | ^ Magnet | ||
+ | | Magnet BZH Number | ||
+ | | Frequency | ||
+ | | Cryobore | ||
+ | | Design Letter | ||
+ | | Design Version | ||
+ | ^ Cryostat | ||
+ | | Cryostat Serial | ||
+ | | Dewar Type | D262 | | ||
+ | | RT Bore | ||
+ | ^ Helium | ||
+ | | He Total Volume | ||
+ | | He Holdtime | ||
+ | | He Boil-Off | ||
+ | ^ Nitrogen | ||
+ | | N2 Total Volume | ||
+ | | N2 Holdtime | ||
+ | | N2 Boil-Off | ||
+ | | N2 Vessel Height | ||
+ | ^ Helium Sensor | ||
+ | | Part Number | ||
+ | | Min. Allowed Level | 46 % | | ||
+ | | Min. Level Energizing | ||
+ | | Active Length | ||
+ | ^ He level graph || | ||
+ | | Point1 Level | 0 % | | ||
+ | | Point1 Volume | ||
+ | | Point1 Level | 37 % | | ||
+ | | Point1 Volume | ||
+ | | Point1 Level | 100 % | | ||
+ | | Point1 Volume | ||
+ | ^ Nitrogen Sensor (if present) | ||
+ | | Part Number | ||
+ | | Min. Allowed Level | % | | ||
+ | | Active Length | ||
+ | i premo el botó < | ||
+ | Premo el botó < | ||
+ | Instal·lo i configuro la darrera versió del MICS (2.09i, juny de 2021), i importo el fitxer BIS de l' | ||
===== PENDENT: Configuració de l' | ===== PENDENT: Configuració de l' | ||
Line 382: | Line 396: | ||
</ | </ | ||
+ | ===== Adreces IP dels mòduls de la consola ===== | ||
+ | |||
+ | <WRAP important> | ||
+ | Aquestes adreces només donen accés als mòduls de la consola si es visiten des de l' | ||
+ | </ | ||
+ | |||
+ | * FIXME [[http:// | ||
+ | * [[http:// | ||
+ | * [[http:// | ||
+ | * [[http:// | ||
+ | * [[http:// | ||
+ | * [[http:// | ||