Company logo
User Documentation BioSpec/PharmaScan with AVANCE NEO, V004SearchContentsHome
 
 
Technical Data BioSpec 117/16 USR
STANDARD CONFIGURATION
 
Magnet type
Ultra Shielded & Refrigerated (USR)
1H resonance frequency
500.3 MHz
Diameter of magnet bore
154 mm
Max. gradient strength
570 mT/m @ 150 A
Max. linear slew rate
5130 T/m/s
Shim set
X, Y, Z, Z2, ZX, ZY, X2-Y2, 2XY, Z3
Transmit/receive (T/R) channel
Two 1H parallel transmit / receive channels.
Transmitter power
1000 W (2x 500 W)
Standard RF coil
transceiver volume coil, ID = 72 mm
OPTIONS
 
Broadband T/R channel
Additional broadband transmit / receive channel.
Parallel receiver
4, 8, 16
Parallel transmitter
up to 8 parallel transmitter
High power gradient upgrade
760 mT/m, 6840 T/m/s @ 200 A
High power transmitter upgrade
up to 2000 W (2x 1000W)
Gradient inserts
none
RF Coils
Volume and Surface Coils,
Phased Array Coils,
X-Nuclei Volume and Surface Coils,
MRI CryoProbe ™
 
INSTALLATION
 
Rating
400/230 VAC, 3W+N+PE, 50/60 Hz, 80 A
Overvoltage category: II
Degree of pollution: 2
IP Classification: 20
Heat dissipation (*)
into air (technical room):
~ 7.5 kW (standard) plus
~ 7 kW High Power Gradient Amplifier (option)
~ 1.2 kW MRI CryoProbe (option)
~ 2.0 kW Parallel Transmit Cabinet (option)

into air (MR system in magnet room): ~ 0.5 kW
into air (operating room): ~ 0.5 kW
into facility cold water supply: ~ 11.5 kW
Acoustic sound pressure
Technical room: typically 70 dB(A)
Magnet room / examination: up to 100 dB(A)
 
(*) The power consumption and the heat dissipation depend on the actual system configuration and on the operation state of the MR instrument. For site planning purpose, see details in the instrument specific Site Planning Information.