For further information concerning the positioning, image settings and acquisition parameters refer to the Accompanying Documentation of your PET/MR-system, or contact the Bruker Service.
When the source is used to assess the maximum sensitivity, the source should be in the center of the FOV so make sure the level is on the orifice marked by the central black circle. This best assed on the position of the source in the image.
Acquisition duration for assessing the sensitivity is 1 minute. For assessing the spatial resolution 10 min is best.
The source is supplied closed/sealed by Eckert & Ziegler. You will find the respective manufacturer certificate in Certificate for Sensitivity and Resolution Phantom. Furthermore, observe the handling instructions supported by Eckert & Ziegler concerning the phantom’s handling.