Removing the Detector Tube from the MR Instrument
Removing requires the following steps:
1.
Remove the RF coil and all samples from the instrument.
2.
Deflate the pneumatic fixation of the detector tube by turning the valve of the Pneumofix in the OPEN position.
3.
Gently pull the detector tube a few cm out and leave it in this position.
4.
Unlock the support shell and rotate it such that the axis of the shell is aligned with the axis of the magnet.
5.
Move the trolley behind the magnet bore and secure it by the wheeler breaks.
6.
Gently lift the PET insert and pull it completely out and onto the support shell.
7.
Secure the PET Insert on the support shell using the safety belts.
8.
Depending on space available, rotate the shell or proceed with the next step prior to rotating.
9.
Open the wheeler breaks and move the PET insert to its storage position.