Height Adjustment
The height of the support of the detector tube has to be adjusted according to the height of the magnet bore of the MR instrument. This is performed by Bruker Service during installation. In case you need re-adjustment, call Bruker Service or proceed as follows:
Remove the detector tube and store it safely for example on the ground floor or a table. Due to the weight of the tube, this needs to be done at least by two persons.
 
CAUTION
Heavy load.
Some Service actions require handling of heavy loads.
Ensure compliance with worker protection regulatory limits.
Potentially handle with at least 2 persons where applicable.
Potentially, hire professional staff trained in handling heavy loads.
 
Secure the detector tube from rolling or falling.
Then partially removal the central allen screw. There is no need to remove this screw completely. By turning the large M35 nut clockwise or counter-clockwise, the height of the detector tube support can be adjusted to a lower or higher height.