Troubleshooting
Status / Indication | Problem / Possible cause | Check / Remedy / Corrective Action |
Tilt the adapter a few times up and down while the water circulation is on. | ||
Bad ECG signal | Check the connections inside the cassette and at the SAI module. | |
Twist the cables (BioSpec, ICON). | ||
Bad respiration signal | Check position of respiration pad and tightness of Luer connections. | |
Unusually high amount of anesthesia gas necessary | Check cassette cover and all connections for tightness. | |
Reduce at the Air Circulation System air flow to activated charcoal container (BioSpec, ICON) | ||
Accumulation of anesthesia gas in the cassette | Increase at the Air Circulation System air flow to activated charcoal container (BioSpec, ICON) | |
Anesthesia mask and tooth bar in mice cassettes are difficult to connect at the rear end | Remove the rear plate first, connect everything, attach the rear plate again to the cassette. | |
Temperature instable (Xtreme) | Possibly interchanged Air In and Air Out hoses. | |
Animal not correctly positioned in field of view (Albira) | Adjust the telescopic arm of the adapter to correct the position for the animal cassette in use. | |
No reception of ECG and temperature signal (Albira) | Check if the correct (labeled "ECG sensor") D-Sub receptacle was used on the Albira interface panel. |