Circularly Polarized RF Coils
This description applies to the following types of RF coils (examples):
Circularly polarized volume coils with active detuning, for example RF RES 400 1H 112/086 QSN TR AD.
Circularly polarized transceiver volume coils, for example RF RES 400 1H 075/040 QSN TR.
 
Although the coil can be tuned and matched inside or outside the magnet bore, it is recommended to tune/match the coil while it is already loaded with the sample and in its final position inside the magnet.
Especially RF coils with active detuning are best to be tuned/matched while the separate receive-only coil and the sample is already in its final position inside the magnet.
 
1.
Connect the RF coil at the corresponding socket of the RF interface
2.
Start the tune/match from ParaVision.
3.
Adjust and optimize the tune and match condition using the tune and match knobs (labeled T or M) of the first channel indicated by yellow rods and labelled “yellow” in the RF Interface Display.
4.
Proceed the same way with the second channel indicated by green rods and labelled “green” in the RF Interface Display.
5.
Stop the tune/match via the RF Interface Display or in ParaVision.