Cleaning and Disinfection
Cleaning
The Animal Cradles and adapters for the different imaging instruments are not watertight. Use dry cleaning whenever possible or dampen a soft, lint-free cloth with water or a diluted cleaning solution of an all-purpose cleaner and wipe the surface carefully. Do not immerse.
 
Disinfection
We recommend to use surface disinfection, for example Microbac® forte.
Disinfection with VHP, a vapor form of hydrogen peroxide (H2O2), is under investigation. Please ask the hotline for available results.
 
 
 
NOTICE
Material damage of the device
Cleaning of the device with the following chemicals will damage the device.
Do not use acetone, ketones, hexanes, acids and alkalis.
Do not use alcohols such as for example ethanol, propanol, or isopropyl alcohol (isopropanol) or products, containing alcohols such as Bacillol®.
Do not use volatile cleaners like thinner or benzine.
 
 
 
NOTICE
Damage of the Device
Sterilization using an Autoclave is not possible. The material of the device will become damaged.
Do not insert the device into an Autoclave for sterilization.