Cleaning and Disinfection
Cleaning
The Animal Cradles and adapters for the different imaging instruments are not watertight. Use dry cleaning whenever possible or dampen a soft, lint-free cloth with water or a diluted cleaning solution of an all-purpose cleaner and wipe the surface carefully. Do not immerse.
Disinfection
We recommend to use surface disinfection, for example Microbac® forte.
Disinfection with VHP, a vapor form of hydrogen peroxide (H2O2), is under investigation. Please ask the hotline for available results.
| |
Material damage of the device Cleaning of the device with the following chemicals will damage the device. - •
- Do not use acetone, ketones, hexanes, acids and alkalis.
- •
- Do not use alcohols such as for example ethanol, propanol, or isopropyl alcohol (isopropanol) or products, containing alcohols such as Bacillol®.
- •
- Do not use volatile cleaners like thinner or benzine.
|
| |
Damage of the Device Sterilization using an Autoclave is not possible. The material of the device will become damaged.
- •
- Do not insert the device into an Autoclave for sterilization.
|