Operational Qualification
Wear and tear or defects may disrupt operational qualification
. Regular checks avoid reduced image quality or acquiring corrupt data. It is useful to log the results of the qualification checks over time to detect trends.
|
| | |
| - ▪
- Verify gradient amplifier matching.
- ▪
- Verify gradient and preemphasis performance.
| |
| - ▪
- Verify functionality and map shim calibration.
| |
RF amplifier (transmitter) performance | - ▪
- Verify transmitter output power.
- ▪
- Verify transmitter linearity.
| |
| - ▪
- Verify receive channel sensitivity and phase adjustment between multiple receiver.
| |
Signal-to-Noise / RF coil performance | - ▪
- Check RF Coils for visual damage.
- ▪
- Perform QA SNR Coil Checks as described in the RF coil documentation using the corresponding setup, procedure, and QA phantom.
| Monthly, recommended every week |
Imaging and spectroscopy performance | - ▪
- Perform image and spectroscopy quality checks as given in the acceptance procedure.
| Every year, recommended every month |
| | After every service action on the gradient amplifier or gradient coil. |