Operational Qualification
Wear and tear or defects may disrupt operational qualification . Regular checks avoid reduced image quality or acquiring corrupt data. It is useful to log the results of the qualification checks over time to detect trends.
 
MR Instrument
Object or Function
What is checked
Interval
Gradient performance
Verify gradient amplifier matching.
Verify gradient and preemphasis performance.
Every year
Shim performance
Verify functionality and map shim calibration.
Every year
RF amplifier (transmitter) performance
Verify transmitter output power.
Verify transmitter linearity.
Every year
Receiver performance
Verify receive channel sensitivity and phase adjustment between multiple receiver.
Every year
Signal-to-Noise / RF coil performance
Check RF Coils for visual damage.
Perform QA SNR Coil Checks as described in the RF coil documentation using the corresponding setup, procedure, and QA phantom.
Monthly,
recommended every week
Imaging and spectroscopy performance
Perform image and spectroscopy quality checks as given in the acceptance procedure.
Every year,
recommended every month
Image Orientation
Check image orientation
After every service action on the gradient amplifier or gradient coil.