|  | Technical Data BioSpec 47/40 USR | | | Ultra Shielded & Refrigerated (USR) | | | | | | | | | | X, Y, Z, Z2, ZX, ZY, X2-Y2, 2XY, Z3 | Transmit/receive (T/R) channel | Two 1H parallel transmit / receive channels. | | | | transceiver volume coil, ID = 197 mm |
| | | Additional broadband transmit / receive channel. | | | | up to 8 parallel transmitter | High power gradient upgrade | 230 mT/m, 650 T/m/s @ 300 A | High power transmitter upgrade | | | | | Volume and Surface Coils, Phased Array Coils, X-Nuclei Volume and Surface Coils, MRI CryoProbe ™ |
| | | 400/230 VAC, 3W+N+PE, 50/60 Hz, 80 A Overvoltage category: II Degree of pollution: 2 IP Classification: 20 | | into air (technical room): ~ 7.5 kW (standard) plus ~ 7 kW High Power Gradient Amplifier (option) ~ 1.2 kW MRI CryoProbe (option) ~ 2.0 kW Parallel Transmit Cabinet (option)
into air (MR system in magnet room): ~ 0.5 kW into air (operating room): ~ 0.5 kW into facility cold water supply: ~ 11.5 kW | | Technical room: typically 70 dB(A) Magnet room / examination: up to 100 dB(A) |
(*) The power consumption and the heat dissipation depend on the actual system configuration and on the operation state of the MR instrument. For site planning purpose, see details in the instrument specific Site Planning Information. |