Normal Operating Conditions
For a safe operation of the instrument, it must be operated under Normal Operating Conditions.   
This includes compliance with environmental conditions (see System Owner Manual Environmental Conditions) as well as:
 
The device is for indoor use only.
Cooling water (and compressed air for PET option) provided and operational.
Quench Pipe installed and operational (not required for cryogen free magnets).
Magnet supervision installed and operational.
All doors and covers of the instrument and the cabinets are closed.
The door of the Faraday cage and/or of the CCM are closed.
RF coils intended for use are connected and loaded either with a dedicated sample or phantom.
The instrument is READY without indication of error messages.
No electrical or protective grounding modifications of the instrument.
No mechanical modifications of the instrument.
 
 