Company logo
User Documentation BioSpec/PharmaScan with AVANCE NEO, V004SearchContentsHome
 
 
Safety
The source for Normalization is only suited for using in PET/MR-systems. The source is supplied closed/sealed by Eckert & Ziegler. You will find the respective manufacturer certificate in Certificate for Normalization Phantom.
 
WARNING
Hazard of injury (also irreversible) caused by radioactive substances.
The phantom contains the radioactive isotope Ge68.
The phantoms may only be used by trained and qualified personal familiar with the applicable national and local radiation and safety regulations.
Always observe all safety-related information given in the accompanying documentation by Eckert & Ziegler.
In compliance with the local and national radiation safety regulations use the following safety equipment as e.g.:
Personal dosimeter.
Personal protective equipment should be worn, which includes lab coat, gloves and safety glasses.
Avoid direct contact with the source. Follow the safety aspects given by the local radiation safety regulations.
 
NOTICE
Property damage caused by improper handling.
Observe the following instructions and all safety-related information to avoid damaging the phantoms.
Avoid any kind of mechanical influence which could damage phantom, like e.g. bumps or shakes.
Before taking the phantom for imaging, ensure the corresponding source is not leaking.
Set-up the acquisition so it is ready to start once the phantom is in position.
Do not leave a source inside the instrument longer than necessary.