Company logo
User Documentation BioSpec/PharmaScan with AVANCE NEO, V004SearchContentsHome
 
 
Cradle Interface “Air”
This type of animal cradle has as interface a connector type “air” so that in addition to anesthesia and animal monitoring, it can be heated via warm water, heat exchanger and air stream (options).
Prerequisites
Body Temperature Conditioning System (option) installed.
Air Supply & Gas Removal Pump    (option) installed.
 
The warm water based Body Temperature Conditioning System creates the heat that is exchanged in an water/air heat exchanger located in the ATS connector prior to the animal cradle. The Air Supply & Gas Removal Pump then circulates the air through the closed cradle.
 
Details on “where to connect what” can be found in the table below.
 
Cradle Connector version "Air". The warm air is supplied via (5) and a diffusor hose (not shown). Removal of air through (6).
 
Interface of the Cradle Connector towards the animal side.
No.
Function
Part that can be connected
1
ECG connectors
SA Instruments Inc. ECG connectors.
2
Temperature Sensor
SA Instruments Inc. Temperature Sensor.
3
Opening
Opening of 17 x 10 mm (W x H) for customized hoses/cables put through.
4
Anesthesia Suction (WAG)
Hose to remove WAG from anesthesia mask.
5
Warm air IN
To connect Warm Air Diffusor:
Mouse P/N T174076
Rat P/N T174077
6
Warm air OUT
Suction for air.
7
Anesthesia Supply
Anesthesia supply hose.
8
Respiratory Pad
To connect SA Instruments respiratory pad with Luer male.