Company logo
User Documentation BioSpec/PharmaScan with AVANCE NEO, V004SearchContentsHome
 
 
Reliability Checks and Tasks
 
PET Instrument
Object or Function
What is checked
Interval
Compressed Air
▪
Check pressure and flow of the compressed air.
Weekly
Air filter, ventilators in electronic cabinet
▪
Check functionality of ventilators.
Monthly
▪
Clean/replace filters.
Every 6 months
Software / Workstation
▪
Check disc space.
▪
Save and archive configuration.
▪
Create and archive backup software image.
Usage dependent
 
 
PET Instrument
Object or Function
What is to be replaced
Interval
Replacement of radioactive QC sources
▪
PET Source NA22/0.37 IM NU4
▪
PET Source GE68/7.40 IM NU4
after 3 years
after 6 years
PET/MR Workplace Computer
▪
To prevent unexpected failure and data loss, exchange the computer hardware.
after 3 years
PET RECO Server, incl. GPU and RAID system
▪
To prevent unexpected failure and data loss, exchange the computer hardware.
after 3 years
PET acquisition hard disk
▪
To prevent unexpected failure and data loss, exchange the computer hardware.
after 3 years
Software Upgrades
▪
Software integrity and bug fixes.
Upgrade with every software release.