Company logo
User Documentation BioSpec/PharmaScan with AVANCE NEO, V004SearchContentsHome
 
 
Reliability Checks and Tasks
 
PET Instrument
Object or Function
What is checked
Interval
Compressed Air
Check pressure and flow of the compressed air.
Weekly
Air filter, ventilators in electronic cabinet
Check functionality of ventilators.
Monthly
Clean/replace filters.
Every 6 months
Software / Workstation
Check disc space.
Save and archive configuration.
Create and archive backup software image.
Usage dependent
 
 
PET Instrument
Object or Function
What is to be replaced
Interval
Replacement of radioactive QC sources
PET Source NA22/0.37 IM NU4
PET Source GE68/7.40 IM NU4
after 3 years
after 6 years
PET/MR Workplace Computer
To prevent unexpected failure and data loss, exchange the computer hardware.
after 3 years
PET RECO Server, incl. GPU and RAID system
To prevent unexpected failure and data loss, exchange the computer hardware.
after 3 years
PET acquisition hard disk
To prevent unexpected failure and data loss, exchange the computer hardware.
after 3 years
Software Upgrades
Software integrity and bug fixes.
Upgrade with every software release.