Accessories
Under normal operating conditions no risk to the operator arises from the usage of Bruker accessories. Accessories that are not tested and released by Bruker for compatibility with the instrument bear a potential risk.
Some accessories require specific handling as listed in the sections below.
 
WARNING
Danger of injury from electrical shock.
1.
Defective devices or devices having insufficient safety standard bear the risk of electrical hazard.
Use only devices and line power cables with approval by nationally recognized testing laboratories (NRTL).
Do not use defective devices or cables or devices showing safety relevant signs of wear and tear.
Check the water hoses and connections before using them in the imaging instrument (where applicable).
Do not use leaking devices in the imaging instruments.
 
 
CAUTION
Risk of personal injury and environmental damage
In case water, for example from the Animal Body Temperature Conditioning water circuit or from the Animal Warming Blankets is spilled and potentially entered the RF Coil (MR instruments), safety measure have to be followed to dry up the instrument in a safe way.
In case of spills other than water e.g. from samples containing chemicals or radioactive tracers, see section Removal of Spilled Sample Fluids or section Radioactive Tracer and Samples where applicable.
Stop running scans and switch the scanner to the stand-by mode.
Dry all wet surfaces and devices of the MR instrument using a dry cloth.
Then, where applicable for RF Coils: Remove the RF Coil from the instrument and allow spilled water to flow out of the coil by placing it in a suitable manner.
Ensure that all wet areas were dried before switching the instrument on again or using the RF Coil. If unclear, let the instrument dry afterwards for an appropriate time, potentially for several hours.
 
Further information