Company logo
User Documentation BioSpec/PharmaScan with AVANCE NEO, V004SearchContentsHome
 
 
Exposed Area
The exposed area is the three-dimensional area in which the magnetic field strength is between 0.1 and 0.5 mT. It is defined by the corresponding stray field lines of the particular magnet types. The exposed area is not a general area of risk.
 
 
NOTICE
Malfunction of laboratory equipment sensitive to magnetic fields
Within the Exposed Area, sensitive equipment may still experience impairment of functionality, disturbance, or defects.