Planned Magnet Maintenance
Planned Maintenance has to be initiated and ordered by the System Owner.
 
Planned Maintenance actions to be performed by Bruker Service at the magnet are:
Object or Function
What is checked
Interval
Cooling Water
Check Dual Loop Heat Exchanger (DLHE) for cold water supply.
Yearly
Magnet He level
Check / refill He.
Yearly (more often in case refrigerator operation was interrupted)
 
Planned Maintenance actions to be performed by Bruker Magnet Service are:
Object or Function
What is checked
Interval
Compressor Adsorber
Compressor Adsorber needs refurbishment (including He refill).
2 years
Magnet Cold Head
Cold head needs refurbishment (including He refill).
4 years
Emergency Magnet Quench
Check of Magnet Emergency Quench button.
2 years
Magnet Supervision
Check BMU and logfiles.
2 years