Replacing BMU Batteries
Procedure   
▪
Unlock drawers containing the batteries and remove batteries.
▪
Insert the new batteries as indicated by the polarity symbols (+ and -) marked.
▪
Push the batteries drawer in the compartment.
▪
Press the BMU battery check button shortly and activate the menu with F1 followed by F9: Batteries are OK as long as no “battery alarm” occurs and the voltage displayed is > 20V.
 
Batteries replacement cases for BMU 4.x:
▪
The 16 batteries must be changed every six months in case of a normal use (BMU has always been connected to the mains).
▪
If the BMU has been disconnected from mains and the batteries have not been removed during that time, you must replace the 16 batteries.
▪
“Battery alarm” occurs and the voltage displayed is < 20V (see last step above).