Replacing BMU Batteries
Procedure   
Unlock drawers containing the batteries and remove batteries.
Insert the new batteries as indicated by the polarity symbols (+ and -) marked.
Push the batteries drawer in the compartment.
Press the BMU battery check button shortly and activate the menu with F1 followed by F9: Batteries are OK as long as no “battery alarm” occurs and the voltage displayed is > 20V.
 
Batteries replacement cases for BMU 4.x:
The 16 batteries must be changed every six months in case of a normal use (BMU has always been connected to the mains).
If the BMU has been disconnected from mains and the batteries have not been removed during that time, you must replace the 16 batteries.
“Battery alarm” occurs and the voltage displayed is < 20V (see last step above).