Troubleshooting
Problem Symptom Message |
Probable Cause | Action Remedy
|
Turning the Tuning/Matching Rods does not tune/match the CryoProbe | Mechanics of Tuning/Matching Rods not functional. |
|
Brightest image area not where expected. | The Reference Power was not correctly set for the desired region. | Readjust the Reference Power. |
Images are too noisy | The spatial resolution is too high. | Reduce matrix, increase FOV, or increase slice thickness. |
The lifting mechanism of the animal bed did not work. |
| |
The distance of the animal’s head to the coil is too large. | Adjust the center of gravity of the animal bed. | |
Image shows only Noise | The RJ45 (LAN) connector at the CCM filter plate was not plugged in. | Plug in the RJ45 (LAN) connector at the CCM filter plate. |
Water condensation or ice in the coil head. | The TCU heater (or the TCU itself) was switched off or automatically switched off during a power outage. |
|
Water condensation or ice at the rear of the CryoProbe where interfaced to the interconnection. | Bottom heater is off. | Please contact the MRI CryoProbe Service. |
The animal bed is locked in the magnet. | The lifting mechanism of the animal bed is stuck in the „up“ position. |
|