Troubleshooting
Problem Symptom Message |
Probable Cause | Action Remedy
|
Automatic Wobble stops with error message. | CryoProbe wobbled the first time after switching the electronics on. | Step width of automatic tuning initially too large: repeat automatic wobble. The second time should work. |
Sweep Width too large for step width of algorithm. | Reduce Sweep Width to 5 or 8 MHz. | |
Default values Tuning % Matching % not adequate. |
| |
PV 360 V3.4 bug: auto wobble does not work and display shows only one curve |
| |
Images are too noisy. | CryoProbe not plugged in. Images are acquired with the TX/RX Volume Coil. | Plug in the white coil connector (ODU). |
Image shows only noise | The RJ45 (LAN) connector at the CCM filter plate was not plugged in. | Plug in the RJ45 (LAN) connector at the CCM filter plate. |
Image brightness inhomogeneous. | Image brightness always differs with distance from the coil elements. However, element image brightness can be scaled relative to the other array element in order to achieve best homogeneity. | See section Quality Assurance, Tips for Array Coils. |
Water condensation or ice in the coil head. | Cooling Unit 4: |
|
| Cooling Unit 5: |
|
Water condensation or ice at the rear of the CryoProbe where interfaced to the interconnection. | Bottom heater is off. | Please contact the MRI CryoProbe Service. |
The animal bed is locked in the magnet. | The lifting mechanism of the animal bed is stuck in the „up“ position. |
|