Starting Wobble Using “Start” (Automatic Adjustment)

For coil elements using varicaps, wobble can be started as automatic adjustment. An underlying algorithm performs the otherwise manual steps of tuning and matching automatically. Its performance can be supervised in the “Acq/Reco Display”.

The resulting tune and match values are stored within the context of the current study. They can also be manually stored for all studies which use the identical coil configuration.

Storing the wobble curve:

When performing wobble as auto-adjustment the parameter Store Wobbleresult on the Result Card can be set to save the wobble curve itself as an adjustment result. The curve can be stored, no matter if the coil can actually be wobbled automatically or not.

The following table summarizes the performed actions when pressing Start:

Coil

Stored Wobbleresult disabled

Stored Wobbleresult enabled

No varicaps

- Nothing is done. The adjustement
  immedately finishes successfully.

- A single wobble curve is recorded
  and stored.

Varicaps

- The coil element is wobbled
  automatically.

- The resulting tune/match values
  are stored.

- The coil element is wobbled
  automatically.

- The final wobble curve is stored
  along with the resulting tune/match
  values.