MRI RF Coil Terminology

MRI RF Coil Terminology

TERM

Description

1H, 13C, 23Na (examples)

proton, carbon-13, sodium-23

1H/X coil

A double-resonant RF coil;
both resonance frequencies can be used simultaneously.

1H-X coil

A single-resonant but tuneable RF coil, that can be tuned in the frequency range from 1H to X.

200 (300, 400, 500, …..)

Designed for a magnetic resonance frequency of 200 MHz (…)

2x2
4x1
8x1 (examples)

Array Coil design with 2 elements in 2 directions
Array Coil design with 4 elements in one directions
Array Coil design with 8 elements typically in a ring design

Active High / Active Low

Active High: A coil that can be actively detuned will be on-resonant while pin driver current flows to the coil.

Active Low: A coil that can be actively detuned will be on-resonant while no current flows or even when disconnected.

Check ParaVision -> Configuration -> RF Coils

AD or AD2

Active Detuneable (sometimes named decoupling instead of detuneable), enables Cross Coil Configuration (optimized for Cross Coil operation with minimum compromise when used for reception).

ARR

Array Coil

CP or CRP

Cryo Coil (CryoProbe™)

part number ….V3

designed for use in AVANCE III HD based instruments,
some may also be used for AVANCE NEO – check manual.

part number ….V4

designed for use in AVANCE NEO based instruments

PC

PET Compatible. RF Coil can be used during PET applications. Typically for these RF Coils, attenuation correction maps are provided in ParaVision. Compare PO.

PO

PET Optimized. RF Coil focuses on PET application with regard to FOV, material (low attenuation) etc. Typically for these RF Coils, attenuation correction maps are provided in ParaVision.

QUAD

Quadrature, synonym for circularly polarized electromagnetic field

QSN/QNS

Quadrature RF coil, polarity south-north / north-south / unspecific.

R., M., BR., HRT (examples)

Rat, Mouse, Brain, Heart

RF

radio frequency

RES

Resonator or Volume Coil

RO

Receive-only coil, no signal excitation possible.

SUC

Surface Coil, single or double resonant

TR

Transceiver coil (transmit and receive coil)

TO

Transmit-only transmitter coils, cannot be used for signal detection.

 

MRI RF Coil Glossary

TERM

Description

Array Coil

A RF coil that consists of several elements, typically arranged as one or two dimensional array or ring structure (short ARR).

Body Coil

A RF coil that allows to image parts of the body of an animal, but not the entire animal.

Brain Coil

A RF coil that allows to image the brain of an animal, but not the entire head.

Cross Coil (configuration)

A coil configuration that consists of separate RF coils for transmission and signal detection.

Head Coil

A RF coil that allows to image the entire head of an animal.

RF Coil

A single coil that is used in a RF coil configuration.

RF Coil  Configuration

A set of one or more RF coils that are used in a MRI experiment.

RF Coil Element

One RF channel of a single or multiple element RF coil (array).

Surface Coil

A RF coil that is placed on the surface and allows to image close to the surface (short: SUC).

Volume Coil

A RF coil that allows to image typically homogeneously a volume.

Whole Body Coil

A RF coil that allows to image the entire body of an animal (excluding the tail).