Specific Handling

Switch between 1H and 19F

The coil can be switched between the frequencies of 1H and 19F by pulling/pushing the rod located between the two rods for tuning.

 

Make sure the coil and setup is fixed sufficiently to allow pushing/pulling the switch.

 

 

Rod extensions

All rods can be extended to enable the operation of the coil while located in the magnet.