Technical Data
|
Mechanical dimension
|
Dout = 112 mm Din = 60 mm L = 545 mm |
|
Dimension of RF structure |
Array Coil: +/- 27 mm in z-direction |
|
Distance to center |
133 mm from front plate |
|
Weight of MRI coil |
approx. 2.7 kg |
|
Typical load |
Rat body, up to 400 g body weight |
|
MRI Nucleus |
proton (1H) |
|
Frequency |
300.3 MHz |
|
Polarization |
circularly polarized transmit Volume Coil, 16-channel receive-only Array Coil in 8 x 2 coil element design |
|
Operation mode(s) |
transceiver Volume Coil, receive-only Array Coil |
|
Active detuning |
yes. Transmit Volume Coil Active High, receive-only Array Coil Active Low. Not to be used in combination with other receive-only coils. |
|
Tune / match |
Transmit Volume Coil: manual tune, manual match. Receive Array Coil: pre-tuned, pre-matched. |
|
Max. RF power limits |
peak power: 1200 W pulse length: 5 ms @ max peak power cw power: 5 W |