Technical Data
|
MRI Nucleus |
proton (1H), carbon (13C) or proton (1H), sodium (23Na) or proton (1H), phosphor (31P) |
|
Frequency |
300.3 MHz (1H) and 75.4 MHz (13C) or 79.4 MHz (23Na) or 121.4 MHz (31P) |
|
Polarization |
linear (1H), linear (13C), linear (23Na), or linear (31P) |
|
Operation mode |
transceiver 1H, transceiver 13C or 23Na or 31P |
|
Active detuning |
no |
|
Tune / match |
manual tune, manual match for 1H and X-nucleus |
|
Max. RF power limits |
1H and X-nucleus: peak power: 400 W (1H), 300 W (13C, 31P, 23Na) pulse length: 5 ms @ max peak power cw power: 2 W |
|
Mechanical dimension
|
Dout = 75 mm Din = 40 mm L = 100 mm |
|
Dimension of RF structure |
Volume coil: +/- 20 mm in z-direction around center mark |
|
Distance to center |
25 mm from front plate |
|
Weight of MRI coil |
approx. 520 g |
|
Typical load |
Mouse body up to 30 g, Rat head up to 400 g body weight |