Connecting the Coil
For AVANCE™ III (or higher) MR instruments, connect the 1H TX/RX MRI CryoProbe™ plug with the red socket labelled 1H High Power for AVANCE™ III or 1H/19F TX/RX for AVANCE™ III HD instruments.
RX-only Array CryoProbes are to be connected to the orange socket labelled 1H Low Power for AVANCE™ III and RX-only for AVANCE™ III HD instruments.
1H/X CryoProbes have two plugs. Connect the 1H channel to the red socket and the X-nucleus channel to the blue socket labelled X High Power for AVANCE™ III or X-BB for AVANCE™ III HD instruments.
The coils will be automatically recognized in ParaVision® 5 and 6.
At AVANCE™ I and II instruments, the MRI CryoProbe™ is permanently wired. Manually select the appropriate coil configuration within ParaVision®.
MRI CryoProbes are highly integrated and technically demanding devices. Inadequate handling may cause damage to the CryoProbe.
- Always connect the CryoProbe to the corresponding socket.
- Do not operate the MR Instrument while the CryoProbe is located in the magnet but not connected to the RF coil socket.
- Do not operate a transmit / receive CryoProbe together with any other RF coil in the scanner.