Line Power Distribution Panel
All components of the MR instrument are supplied with power via the Line Power Distribution panel (LPD) installed in the technical room. It provides distinguished connectors for all cabinets and accessories. No user interaction is required.
A central power interruption switch is located at the front side of the panel. Use the switch only during service activity.
In case of emergency, use the rotary switch at the Electric Emergency to switch off the instrument.
 
 
NOTICE
Connecting power consumers other than the MR instrument to the Line Power Distribution panel or connecting instrument components to other sockets than those provided by the Line Power Distribution panel can lead to ground loops and unstable instrument operation.
Do not connect other devices to the Line Power Distribution panel.
Do not modify the connections at the Line Power Distribution panel.
Do not connect instrument components to other sockets, than those of the Line Power Distribution panel.