Magnet mounted Volume Coils
These RF Volume Coils are mounted directly inside the gradient bore and fixed by inflating the pneumatic fixation system.

Pos. No | Description |
|---|---|
1 | Coil housing |
2a | Valve |
2b | Balloon |
2c | Pneumatic rings |
Mounting instructions:
- Insert the coil from the service end and co-align the coil housing (1) with the end of the gradient tube.
- Switch the valve (2a) in the “pump” position.
- Gently pump and inflate the rings (2c) of the pneumatic fixation system using the balloon (2b).
- Connect the coil with the corresponding RF interface socket according to the color labels.
To remove the coil from the magnet, proceed as follows:
- Stop any running scan prior to disconnecting or dismounting a RF coil.
- Unplug the coil using the leverage effect of the coil socket cover.
- Open the valve of the pneumatic fixation system.
- Remove the coil out from the gradient bore.
NOTICE
The components of the pneumatic fixation system can easily be damaged.
- Never pump and inflate the rings of the pneumatic fixation system outside a surrounding bore to avoid exceeded inflation and damage of the system.
- Do not apply high pressure to the components of the pneumatic fixation system. Gently pump by hand.