informatica:hp_z230_400sb_topspin364-post [2021/09/04 19:34] miquel [PENDENT - Error setting parameter "LIFTMODE"] |
informatica:hp_z230_400sb_topspin364-post [2021/09/08 10:15] miquel [PENDENT - Error setting parameter "LIFTMODE"] |
||
---|---|---|---|
Line 622: | Line 622: | ||
<WRAP todo> | <WRAP todo> | ||
- | He enviat a Bruker aquest error i estic a l' | + | * He enviat a Bruker aquest error i estic a l' |
+ | * La resposta és que cal actualitzar el firmware de la placa ELCB | ||
+ | * Actualitzo el firmware però el missatge d' | ||
- | Sortosament, | + | Sortosament, |
</ | </ | ||
Line 652: | Line 654: | ||
at java.lang.Thread.run(Thread.java: | at java.lang.Thread.run(Thread.java: | ||
</ | </ | ||
+ | |||
+ | Segons Bruker, aquest error es resol actualitzant el firmware de la placa ELCBñ. La versió actualment instal·lada és, | ||
+ | |||
+ | ^ Unit ^ Version | ||
+ | | ELCB Firmware | ||
+ | | Bootloader Firmware | ||
+ | |||
+ | i a la pàgina "BSMS Service Web - Setup" ho confirma, | ||
+ | |||
+ | Firmware Status | ||
+ | ^ Unit Name | ||
+ | | BSMS/2 ELCB | unknown | ||
+ | | BSMS/2 SCB20 | ||
+ | | BSMS/2 SLCB/ | ||
+ | |||
+ | Clico a l' | ||
+ | |||
+ | * elcb_firmware_210212.gz | ||
+ | |||
+ | Un cop finalitzada la instal·ació del nou firmware. torno a comprovar la versió instal·lada, | ||
+ | |||
+ | ^ Unit ^ Version | ||
+ | | ELCB Firmware | ||
+ | | Bootloader Firmware | ||
+ | |||
+ | i a la pàgina "BSMS Service Web - Setup" la columna '' | ||
+ | |||
+ | Llavors reinicio TopSpin per comprovar si ja no apareix l' | ||
+ | |||
+ | |||