Exchanging Gradient Inserts
 
CAUTION
There is a risk of personal injury and/or property damage when lifting, carrying or installing the gradient insert.
Due to its weight (≥ 20 kg) the gradient insert has to be handled by at least 2 persons to avoid personal injury or damages.
Ensure compliance with worker protection regulatory limits.
Potentially, hire professional staff trained in handling heavy loads.
 
 
In order to install a gradient insert in the MR instrument, proceed as follows:
1.
Switch the MR instrument in the operation mode System OFF.
2.
Remove all RF coils and animal beds from the system gradient.
3.
Move the insert from the service end of the magnet to the iso-center of magnet.
Information about the distances are labeled on the type plates of the magnet and gradient insert.
4.
Adjust the rotation of the gradient insert in such a manner, that the thick mark is oriented along the physical +Y axis of the magnet coordinate system (12 o’clock position). Refer to section Physical Coordinate System.
5.
Fix the gradient insert with the non-magnetic tools. Therefore, carefully turn the eccentric screws until the gradient insert is well fixed. Do not apply brute force!
The tool is in the scope of delivery.
6.
Disconnect the gradient current connector (Schaltbau) of the system gradient set inside the CCM and connect the equivalent connector of the new insert.
7.
Disconnect the shim current connector (Hypertac) of the main shim set inside the CCM and connect the equivalent connector of the insert.
8.
Feed the water cooling tubes of the insert through the RF shield on the side of the CCM.
9.
Remove the U-shaped bypass tube and connect the cooling tube of the gradient insert at the water distribution panel
10.
Switch the MR instrument in the operation mode System ON.
 
 
NOTICE
Each kind of mechanical force can cause property damage to the gradient shim system and its connectors.
Avoid each kind of mechanical force such as bumps or shakes when connecting the gradients and shims.
Do not twist any housing, sealing or quick-lock connectors.
Always use the protective cover of the shim connector, when the insert or the main shim set is not connected.