Exchanging Gradient Inserts
In order to install a gradient insert in the MR instrument, proceed as follows:
- 1.
- Switch the MR instrument in the operation mode System OFF.
- 2.
- Remove all RF coils and animal beds from the system gradient.
- 3.
- Move the insert from the service end of the magnet to the iso-center of magnet.
- ⇨
- Information about the distances are labeled on the type plates of the magnet and gradient insert.
- 4.
- Adjust the rotation of the gradient insert in such a manner, that the thick mark is oriented along the physical +Y axis of the magnet coordinate system (12 o’clock position). Refer to section Physical Coordinate System.
- 5.
- Fix the gradient insert with the non-magnetic tools. Therefore, carefully turn the eccentric screws until the gradient insert is well fixed. Do not apply brute force!
- ⇨
- The tool is in the scope of delivery.
- 6.
- Disconnect the gradient current connector (Schaltbau) of the system gradient set inside the CCM and connect the equivalent connector of the new insert.
- 7.
- Disconnect the shim current connector (Hypertac) of the main shim set inside the CCM and connect the equivalent connector of the insert.
- 8.
- Feed the water cooling tubes of the insert through the RF shield on the side of the CCM.
- 9.
- Remove the U-shaped bypass tube and connect the cooling tube of the gradient insert at the water distribution panel
- 10.
- Switch the MR instrument in the operation mode System ON.