Company logo
User Documentation BioSpec/PharmaScan with AVANCE NEO, V004SearchContentsHome
 
 
Operational Qualification
Operational Qualification (sometimes also called quality assurance) ensures reproducible performance of the PET instrument in specification. If conducted on a regular basis, there is a high probability to detect degradation of defects of the PET instrument in an early phase and to avoid the acquisition of unqualified PET data. For this purpose it is useful to log the results of the operational qualification results over time.
 
PET Instrument
Object or Function
What is checked
Interval
ATS Center Setting in PET, MR or CT Center
(not applicable for PET Insert).
Check that the center setting of the transport system corresponds to the center of the FOV of the corresponding modality.
Identify the trend over the year and recalibrate if required.
Monthly
PET/MR
PET/CT
co-registration
Check that the co-registration of PET with the MR or CT images is correct and precise.
Identify the trend over the year and recalibrate if required.
Monthly
Image Quality
Check image quantification.
Check energy resolution.
Check image deformation (decompression).
Check image uniformity.
Identify the trend over the year and recalibrate if required.
Weekly
Detector Normalization
Check energy peak position.
Check singles rates.
Check coincidence rates.
Check detector event map.
Every 3 months
Sensitivity
Identify the trend over the year.
Monthly
Spatial Resolution
Use and check the log files of monthly performance qualification tests.
Identify the trend over the year.
Monthly
Gating
Use an external gating generator to acquire and check a gated acquisition.
Every 6 months
 
Further information
How to install Quality Control Samples
How to verify the Alignment of PET and MRI
How to verify the Image Quality
How to verify the Sensitivity
How to verify the Resolution
How to verify the Performance